MAESTRO-DF for Agilent LCMS
- Fully regenerative PSA technology ; reduced risk of gas contamination and phthalate free
- HMI touch screen technology to display the process in real time, including process variables
- Integral oil free compressor in option : fully secure supply
- Quiet thanks to the Soundproofed compressor box and anti-vibration features
- Auto start
- Audible and Alarm display with help menu and history log
- Visual maintenance indication with alarm and history log
- Outlet flow indicator
- Energy saving Mode : Enables the compressor to switch off when nitrogen supply is not required
- Remote access to screen using internet via USB port.
- Fitted with wheels
Model |
Built in compressor |
Outlet Flow rate |
Outlet Pressure |
Dimensions |
Weight |
Maestro-DF |
Yes |
35 L/min N2 @ 99% for Drying, sheath/nebulizing gas |
7 bar (101 psi) |
83 x 43 x 80 cm (33″ x 17″ x 31″) |
118 / 260 |
200 ml/min N2 @ 99,999% |
3 bar (43 psi) |
||||
Technical information |
|||||
Ambient Temp range |
5 – 25°C (41 – 77°F) |
||||
Air Inlet Quality Requirement (units without compressor) |
Dewpoint : -40°C (-40°F) |
||||
Particulate : < 1 micron |
|||||
Oil : < 0.01 mg/m3 |
|||||
Electrical Supply |
220 V ac / 1 ph / 50 Hz |
||||
110 V ac / 1 ph / 50 – 60 Hz |
|||||
Inlet/Outlet connections |
G 1/4” (BSP) Female |
Application | All LC-MS analyzers do not share the same requirements in terms of flow rate, pressure and purity of the gases needed for their operation. F-DGS offer several options for high flow nitrogen requirements. |
---|
All LC-MS analyzers do not share the same requirements in terms of flow rate, pressure and purity of the gases needed for their operation. F-DGS offer several options for high flow nitrogen requirements.