007-624
- Bonded and crosslinked
- Specifically designed for volatile organic compound analyses
- Similar phases: DB-624, DB-1301, Rtx-502, DB-608, VOCOL, RTX-Volatiles, PB-608, Rtx-35, AT-624, BP-624, CPSil-13CB
- USP Phase Specification: G43
EPA Method applications: volatiles and semi-volatiles in drinking water, ground water, pesticides, herbicides, etc.
007-624 offer three variations of cyanopropylphenyl substitutions on a methylpolysiloxane backbone. These phases have been specifically designed for the analysis of volatile organic compounds and pesticides as specified in many EPA Methods including 502.1, 502.2, 503.1, 524.1, 524.2, 601, 602, 608, 624, 8010, 8011, 8015, 8020, 8240, and 8260. By using 007-624 phase types on 0.53mm I.D. columns, the entire range of volatile compounds, including gases, can be analyzed beginning at ambient temperatures.
007-624 Cyanopropylphenyl Methylpolysiloxane columns are most commonly requested in the following column I.D.'s and column lengths. The film thickness ranges are noted in the chart below.
Specifications:
I.D. | 30M. | 50M. | 60M. | 75M. | 100M. | 105M. | STANDARD FILM THICKNESS RANGE | ||
---|---|---|---|---|---|---|---|---|---|
0.25mm | 1.0 - 3.0µm | ||||||||
0.32mm | 1.0 - 5.0µm | ||||||||
0.53mm | 0.5 - 5.0µm | ||||||||
Inquire with tech@quadrexcorp.com about custom lengths and films. |
007-608, 007-624, and 007-502 OPERATING TEMPERATURE RANGE | |
---|---|
film thickness | temperature range |
up to 1.0µm | -50° C to 325°C |
2.0 to 5.0µm | up to 280°C |
BONDED PHASE RETENTION INDEX | |||||
---|---|---|---|---|---|
benzene | 1-nitropropane | 2-pentanone | pyridine | butanol | |
684.05 | 881.36 | 724.27 | 793.53 | 707.78 |
Hot tip:
Don't see a length or film thickness that you require?
Contact us - we do custom configurations all the time!